Sets the result update time. That is the maximum rate in which the power sensor can
output measurement results.
Parameters:
<update_time> Range: 0.0 to 10.0
*RST: 0.1
Default unit: Seconds
SYSTem:SERRor?
Queries the next static error, if available.
Static errors occur when you select conflicting settings. For example in timeslot mode,
a static error occurs with the following settings:
Width of a timeslot: 100 µs
Exclude time at the start of the slot: 40 µs
Exclude time at the end of the slot: 60 µs
Then there is "nothing left" to be measured, and a static error appears.
Static errors, as a rule, prevent the execution of normal measurements.
Errors in SCPI communication are queried using SYSTem:ERRor[:NEXT]?.
Usage:
Query only
SYSTem:SERRor:LIST:ALL?
Queries all changes in the static error queue that have not been read yet and removes
them from the queue.
Example:
SYST:SERR:LIST:ALL?
Response: 0,"reported at uptime:2942; notice;
auto-averaging exceeded maximum time;
Notification",0,"removed at uptime:2944;
notice; auto-averaging exceeded maximum time;
Notification".
Usage: Query only
SYSTem:SERRor:LIST[:NEXT]?
Queries the list of static error changes for the oldest entry and removes it from the
queue.
Returns an error number and a short description of the error.