Remote control commands
R&S
®
NRPxxS(N)
92User Manual 1177.5079.02 ─ 15
Accelerating measurements
Using [SENSe<Sensor>:][POWer:][AVG:]FAST ON, you can accelerate the mea-
surement as follows:
●
Chopper is disabled.
●
Average count is set to 1, no matter which average count you have set.
Thus, the overall measurement time is only defined by the aperture time, and the mea-
surement time for a fast measurement is calculated as follows:
MT = APER
The fast measurement setting delivers up to 100 000 measurements per second with-
out any blind time over randomly long time periods. Programming examples are given
in Chapter 10.2, "Performing the fastest measurement in continuous average mode",
on page 143.
Remote commands:
[SENSe<Sensor>:][POWer:][AVG:]APERture......................................................................92
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:CLEar...............................................................92
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:COUNt?............................................................93
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:DATA?.............................................................. 93
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE.................................................................93
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:STATe...............................................................93
[SENSe<Sensor>:][POWer:][AVG:]FAST............................................................................ 94
[SENSe<Sensor>:][POWer:][AVG:]SMOothing:STATe.......................................................... 94
[SENSe<Sensor>:][POWer:][AVG:]APERture <integration_time>
Sets the duration of the sampling window in continuous average mode. During this
time interval, the average signal power is measured.
The minimum value is implemented for fast unchopped continuous average measure-
ments. See also Chapter 10.2.2, "Triggered fast unchopped continuous average mea-
surement", on page 145.
Parameters:
<integration_time> Range: 8.0e-6 to 2.00
*RST: 0.02
Default unit: Seconds
Manual operation: See "Aperture Time" on page 46
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:CLEar
Clears the contents of the result buffer in continuous average mode.
Example:
BUFF:CLE
Usage: Event