Page #48 background imageLoading...
Page #48 background image
Browser-based user interface
R&S
®
NRPxxS(N)
48User Manual 1177.5079.02 ─ 15
Number of Timeslots..................................................................................................... 48
Nominal Width...............................................................................................................48
Start Exclude.................................................................................................................48
End Exclude..................................................................................................................48
Number of Timeslots
Sets the number of simultaneously measured timeslots. Up to eight slots can be
selected.
Remote command:
[SENSe<Sensor>:][POWer:]TSLot[:AVG]:COUNt
Nominal Width
Sets the length of a timeslot in seconds.
Remote command:
[SENSe<Sensor>:][POWer:]TSLot[:AVG]:WIDTh
Start Exclude
Sets a time that is to be excluded at the beginning of the measurement period.
Remote command:
[SENSe<Sensor>:]TIMing:EXCLude:STARt
End Exclude
Sets a time that is to be excluded at the end of the measurement period.
Remote command:
[SENSe<Sensor>:]TIMing:EXCLude:STOP
6.4.4 Trace mode
Describes the parameters of the trace measurement.
Further information:
Chapter 9.7.4, "Trace measurement", on page 98
Detailed description of the trace measurement mode and its remote commands
Access: main dialog of the web user interface > navigation pane > "Trace"
Measurement modes

Table of Contents

Other manuals for R&S NRP Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S NRP Series and is the answer not in the manual?

R&S NRP Series Specifications

General IconGeneral
Product TypePower Sensor
Connector TypeSMA, 3.5 mm, 2.4 mm
InterfaceUSB, LAN
Storage Temperature-40°C to +70 °C
DimensionsDependent on model
WeightDependent on model
CompatibilityR&S Instruments
TypeDiode

Related product manuals