Remote control commands
R&S
®
NRPxxS(N)
140User Manual 1177.5079.02 ─ 15
Controlling the ENABle part.................................................................................. 141
Controlling the negative transition part..................................................................141
Controlling the positive transition part...................................................................142
9.13.1 General status register commands
STATus:PRESet.............................................................................................................140
STATus:QUEue[:NEXT]?................................................................................................ 140
STATus:PRESet
Resets the edge detectors and ENABle parts of all registers to a defined value.
Usage:
Event
STATus:QUEue[:NEXT]?
Queries the most recent error queue entry and deletes it.
Positive error numbers indicate sensor specific errors. Negative error numbers are
error messages defined by SCPI.
If the error queue is empty, the error number 0, "No error", is returned.
Usage:
Query only
9.13.2 Reading the CONDition part
STATus:DEVice:CONDition?
STATus:OPERation:CALibrating:CONDition?
STATus:OPERation:CONDition?
STATus:OPERation:LLFail:CONDition?
STATus:OPERation:MEASuring:CONDition?
STATus:OPERation:SENSe:CONDition?
STATus:OPERation:TRIGger:CONDition?
STATus:OPERation:ULFail:CONDition?
STATus:QUEStionable:CALibration:CONDition?
STATus:QUEStionable:CONDition?
STATus:QUEStionable:POWer:CONDition?
STATus:QUEStionable:WINDow:CONDition?
Usage:
Query only
9.13.3 Reading the EVENt part
STATus:DEVice[:EVENt]?
STATus:OPERation:CALibrating[:SUMMary][:EVENt]?
STATus:OPERation[:EVENt]?
Using the status register

Table of Contents

Other manuals for R&S NRP Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S NRP Series and is the answer not in the manual?

R&S NRP Series Specifications

General IconGeneral
Product TypePower Sensor
Connector TypeSMA, 3.5 mm, 2.4 mm
InterfaceUSB, LAN
Storage Temperature-40°C to +70 °C
DimensionsDependent on model
WeightDependent on model
CompatibilityR&S Instruments
TypeDiode

Related product manuals