Remote control commands
R&S
®
NRPxxS(N)
106User Manual 1177.5079.02 ─ 15
MOVing
Outputs intermediate values to facilitate early detection of
changes in the measured quantity. In the settled state, that
means when the number of measurements specified by the
average count has been performed, a moving average is output.
REPeat
Specifies that a measurement result is not output until the entire
measurement has been completed. This means that the number
of measurement cycle repetitions is equal to the set average
count. If the average count is large, the measurement time can
be very long.
The average count is set using [SENSe<Sensor>:]AVERage:
COUNt on page 103.
*RST:
REPeat
Example:
AVER:TCON REP
Manual operation: See "Filter Terminal Control" on page 51
[SENSe<Sensor>:]AVERage[:STATe] <state>
Available in continuous average, burst average and timeslot mode.
Enables or disables the averaging filter.
Parameters:
<state> *RST: ON
Manual operation: See "<Mode>" on page 45
9.8.2 Setting the frequency
The frequency of the signal to be measured is not automatically determined. For ach-
ieving better accuracy, the carrier frequency of the applied signal must be set.
[SENSe<Sensor>:]FREQuency <frequency>
Transfers the carrier frequency of the RF signal to be measured. This frequency is
used for the frequency-response correction of the measurement result.
The center frequency is set for broadband signals, e.g. spread-spectrum signals, multi-
carrier signals.
Parameters:
<frequency> Range: 0.0 to 110.0e9
*RST: 50.0e6
Default unit: Frequency
Example:
FREQ 10000
Manual operation: See "Frequency" on page 44
Configuring basic measurement parameters

Table of Contents

Other manuals for R&S NRP Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S NRP Series and is the answer not in the manual?

R&S NRP Series Specifications

General IconGeneral
Product TypePower Sensor
Connector TypeSMA, 3.5 mm, 2.4 mm
InterfaceUSB, LAN
Storage Temperature-40°C to +70 °C
DimensionsDependent on model
WeightDependent on model
CompatibilityR&S Instruments
TypeDiode

Related product manuals