Examination
Digital Radiography
AXIOM Iconos R200 AXD3-340.620.13.01.02 41 / 52
Selection
Press this key.
Press this key.
DR scanning is selected.
The display shows 3 scanning programs with 2 directions each.
Select an organ program for DR scanning at the integrated generator console.
Deselection
Press this key or select a different exposure system.
DR scanning is deselected.
Image transfer
Send the images from the imaging system to the LEONARDO workstation us-
ing the "DICOM Send” function.
Releasing an exposure
Release the exposure with the exposure release button in the system remote
control console or via the footswitch in the control room, or table-side (option).
Radiation is released as long as you continue to press the exposure release
button.
The signal lamps for radiation must light up.
Radiation interruption If you let go of the exposure release button during radiation, the exposure is in-
terrupted immediately.
Exposure end Letting go of the exposure release button, a full image memory or reaching the
maximum scene time ends the exposure series.
DR
DR
DR
DR
Warning
If these radiation ON indicators light up although no switch has been actuated
and the current fluoroscopic image cannot be seen on the monitor:
Switch the system off immediately!
Notify SIEMENS Uptime Service.
Discontinue operation of the system.