Examination
Digital Radiography
46 / 52 AXD3-340.620.13.01.02 Operator Manual
Organ programs
All parameters necessary for the exposure are stored in organ programs.
➩ For further details on the organ programs and their programming refer to the
FLUOROSPOT Compact Operator Manual and the POLYDOROS SX register.
The programs are intended for the following examinations:
❏ Program A: Survey images of the urinary tract (kidney and bladder)
with scanning program “Abdomen” and direction of scanning “headwards”
(8 cm/sec scanning speed)
❏ Program B: Orthopedic longleg images with scanning program “Longleg” and
direction of scanning “footwards” (4 cm/sec scanning speed)
❏ Program C: a.p. and lateral images of the spine with scanning program
“Spine” and direction of scanning “headwards” (4 cm/sec scanning speed)
❏ Program D: Survey image of the colon with scanning program “Abdomen”
and direction of scanning “headwards” (8 cm/sec scanning speed)
Program suggestions for FLUOROSPOT Compact
Organ program A B C D
Program name Kidneys Legs Spine ap/lat Colon
Fluoroscopic
mode
Fluoro 1 Fluoro 1 Fluoro 1 Fluoro 1
Supervision//
Carevision
SV // 3 P/s SV // 3 P/s SV // 3 P/s SV // 3 P/s
SDM dominant l, r l, r c c
Dose level 100 100 100 100
Acquisition data
from fluoro
Yes Yes Yes Yes
kV auto auto auto auto
Characteristic
curve (C)
C 16 C 16 C 18 C 21
kV dose reduc-
tion [kV]
133 133 133 133
Focus Large Large Large Large
Regulation stop Yes No No Yes
Max. pulse width
[ms]
30 30 30 30
Film density cor-
rection
0.0 0.0 0.0 0.0
Grid In In
1
No In
Acquisition filter:
auto
No No No No