Examination
Digital Radiography
AXIOM Iconos R200 AXD3-340.620.13.01.02 47 / 52
Peristepping examination sequence for survey images
Take Peristepping exposures as described under Peristepping.
Select an exposure for each exposure position (e.g. a “max. opac image”) of
this series and send it from the imaging system to the LEONARDO worksta-
tion using the “DICOM Send” function.
Filter type
Cu[mm]
0.0 0.2
2
0.2
2
0.0
Acquisition mode Native series Native series Native series Native series
Mark image Yes Yes Yes Yes
Auto collimator Yes Yes Yes Yes
Bones White Black White
3
Black White
4
Black White Black
Contrast (C) 700 700 700 700 700 700 700 700
Brightness (B) 306 450 276 480 276 480 316 440
Scene length[s] 10 30 30 10
Frame rate 13223
Frame rate 23223
Frame rate 33223
Edge filter [%]15151515
Kernel size5555
Harmonization
[%]
30 60 60 30
Harmonization
kernel
127 127 127 127
1
Drive grid out for slender patients
2
Use 0.0 mm copper filter for obese patients
3
When a white display is required for the survey images on the LEONARDO workstation, acquire the images with black display in the
imaging system. Compose these on the workstation and then invert them from black to white at the workstation. This also applies for
one-sided leg displays
4
When a white display is required for the survey images on the LEONARDO workstation, acquire the images with black display in the
imaging system. Compose these on the workstation and then invert them from black to white at the workstation
Organ program A B C D