Accessories
Optional Accessories
20 / 54 AXD3-340.620.13.01.02 Operator Manual
◆ Press both tension levers (4) on the belt holder inward.
◆ Pull the belt holder backward off the accessory rail.
◆ Lift the belt holder over the patient.
◆ Roll up the compression belt halfway using the cap screw.
◆ Place the belt holder next to the tensioning roller.
◆ Press both levers (2) inward on the tensioning roller.
◆ Pull the tensioning roller from the front accessory rail.
◆ Assist the patient in stepping down from the patient table.
◆ Using the cap screw (3), roll up the belt on the tensioning roller.
Radiation protection
value:
The attenuation equivalent of the compression belt is ≤ 0.8 mm Al.
44
22
1
3