Press both tension levers (4) on the belt holder inward.
Pull the belt holder backward off the accessory rail.
Lift the belt holder over the patient.
Roll up the compression belt halfway using the cap screw.
Place the belt holder next to the tensioning roller.
Press both levers (2) inward on the tensioning roller.
Pull the tensioning roller from the front accessory rail.
Assist the patient in stepping down from the patient table.
Using the cap screw (3), roll up the belt on the tensioning roller.
Radiation protection
value:
The attenuation equivalent of the compression belt is 0.8 mm Al.