Accessories
Optional Accessories
AXIOM Iconos R200 AXD3-340.620.13.01.02 19 / 54
Press the tension lever (4) outward.
Check that the holder is seated firmly by pulling on it.
Turn the cap screws (3) on the tensioning roller toward the right.
The belt will tighten.
Push and pull back on the ratchet handle (1), until the belt tension is optimal.
Check the belt tension on the patient once again.
Loosening the
compression belt
Move the patient table to the horizontal position.
Press the ratchet handle (1) back.
Turn the cap screw (3) on the tensioning roller toward the left.
The compression belt will loosen.
Using the ratchet handle, set a lower belt tension.
Removal Move the patient table to the horizontal position.
Press the ratchet handle (1) back.
Turn the cap screw (3) on the tensioning roller toward the left.
The compression belt will loosen.
4
4
3
1