Accessories
Optional Accessories
AXIOM Iconos R200 AXD3-340.620.13.01.02 45 / 54
Changing the cone
shape
Removing
Hold the cone arm with one hand.
Grasp the cone with your other hand and push it out to the side from the hold-
er.
Inserting
Place the cone arm with the opening upwards on a firm surface.
Insert the cone laterally and press it in until it snaps into place.
Attachment Move the patient table to the horizontal position.
The compression carriage should be in the uppermost position.
Hold the compression cone so that it is horizontal.
Insert the pins of the cone arm into the opening in the compression carriage
up to the stop.
The locking lever (1) locks into place.
Pull on the cone arm and check for firm seating of the cone arm.
Position the patient.
Move the cone into the compression position.
Removal Move the patient table to the horizontal position.
Move the compression carriage one notch upward.
The cone must no longer touch the patient.
Press the locking lever (1) and hold it until the lock is released.
Pull the cone arm out completely.
Move the compression carriage into the upper end position.
(1)
For safety reasons, the maximum compression cannot be exceeded. The com-
pression drive is automatically shut off when it reaches the maximum compres-
sion force.