System Overview
Operating and Display Elements
42 / 44 AXD3-340.620.13.01.02 Operator Manual
◆ Loosen (single turn) the two heat shield screws (1) on the lamp.
◆ Remove the heat shield (2).
◆ Remove the defective bulb (2).
Do not touch the bulb with your bare fingers - it can be very hot!
◆ Insert the new bulb using a cloth and make sure both contacts (3) are touching
the frame stop position.
Do not touch the bulb with your bare fingers!
◆ Insert the heat shield to the stop position and tighten the screws (1).
◆ Attach the lamp cover.
Heat shield might be very hot! Please let it cool down before removing it.
(1)
(2)
(3)
Caution
Lamp breaks
Patient and/or user injured by glass splinters
◆ Use only OEM Siemens spare parts