System Overview
Operating and Display Elements
42 / 44 AXD3-340.620.13.01.02 Operator Manual
Loosen (single turn) the two heat shield screws (1) on the lamp.
Remove the heat shield (2).
Remove the defective bulb (2).
Do not touch the bulb with your bare fingers - it can be very hot!
Insert the new bulb using a cloth and make sure both contacts (3) are touching
the frame stop position.
Do not touch the bulb with your bare fingers!
Insert the heat shield to the stop position and tighten the screws (1).
Attach the lamp cover.
Heat shield might be very hot! Please let it cool down before removing it.
(1)
(2)
(3)
Caution
Lamp breaks
Patient and/or user injured by glass splinters
Use only OEM Siemens spare parts