How to fix 'Memory Full!' error on Siemens Measuring Instruments?
D
David SteeleAug 18, 2025
If the Siemens Measuring Instruments displays 'Memory Full!', it means you've tried saving site data when the memory is full. To fix this, delete an old site or clear the Datalogger memory to free up space for new data.
A
Amanda PadillaAug 20, 2025
What to do if Siemens Measuring Instruments displays '- - -F- - - Fault Alarm'?
B
butlerchristopherAug 20, 2025
If the Siemens Measuring Instruments shows '- - -F- - - Fault Alarm', it could be due to a loss of signal strength (ALC) or a change in Rx signal location (Beam Blowing). Try recoupling sensors with fresh couplant or install sensors in Direct mount mode. If the problem continues, contact technical support.
D
Dana TownsendAug 22, 2025
Why Siemens Measuring Instruments displays 'Low Signal - Press '?
H
Helen EstradaAug 23, 2025
If Siemens Measuring Instruments displays 'Low Signal - Press ', it means the receive signal level is insufficient for proper operation during the Initial Makeup. Possible causes include:
* Invoking [Install Complete?] on an empty pipe.
* Insufficient coupling compound.
* A disconnected or broken Sensor cable.
* The pipe needs conditioning.
* Large air bubbles.
* Defective Sensor cables or incorrect connections.
* Incorrect Set Empty routine.
Correct the condition and press to resume the installation. Otherwise, press the to abort and investigate.
J
James WilliamsAug 26, 2025
What causes 'Re-space Index' error on Siemens Measuring Instruments?
J
Jessica WilsonAug 26, 2025
If the Siemens Measuring Instruments displays 'Re-space Index', it means the measured liquid sonic velocity (Vs) is more than +/- 25% of the average Vs range. Check the pipe dimensions and liquid data entries, ensure the correct Sensor Size is entered in the [Install Sensor] menu, and confirm the sensor spacing in the [Install Sensor] menu.
J
John DeleonAug 27, 2025
What does 'Invalid Setup (use Direct Mode)' mean on Siemens Measuring Instruments?
N
nelsonandreaAug 28, 2025
If Siemens Measuring Instruments displays 'Invalid Setup (use Direct Mode)', it indicates invalid sensor spacing, liquid or pipe parameters, preventing Initial Makeup completion. This could be due to out-of-range data, invalid conditions like overlapping Sensors in Reflect Mode, or other factors. If selecting Direct Mode doesn't resolve it, review all site setup and Sensor installation choices, especially data entered for pipe and liquid. In Reflect Mode, if the pipe wall signal may impinge upon the liquid signal, use Direct Mode instead. Press , , , or to abort and program other site data. Contact technical support if needed.
R
Richard MurphyAug 30, 2025
What does 'Detection Fault' mean on Siemens Measuring Instruments and how to fix it?
T
Thomas MillerAug 30, 2025
If Siemens Measuring Instruments displays 'Detection Fault', it means the pipe and/or liquid conditions do not permit a receive signal that meets the flow detection standards, preventing system operation. To resolve this, try reinstalling the Sensors at a different spacing offset or location on the pipe. Switching from Reflect to Direct Mount may also solve the problem. However, operation may not be possible if there is poor liquid or pipe wall sonic conductivity.
D
david78Sep 1, 2025
Why does Siemens Measuring Instruments say 'Security'?
L
lorikimSep 1, 2025
If the Siemens Measuring Instruments displays 'Security', this means you're changing previously entered data when the security switch is in the [Disable] position or a security code has been entered. To resolve this, change the switch position to [Enable] or enter the previously set security code.
J
Joshua HerreraSep 3, 2025
How to resolve 'Chan Not Setup' on Siemens Measuring Instruments?
M
Mitchell HawkinsSep 3, 2025
If the Siemens Measuring Instruments displays 'Chan Not Setup', it means you're trying to do something that needs a channel to be on. Turn on the channel in [Channel Setup - Channel Enable - Yes]. Remember, you need to finish an "Install" before you can turn on a channel.
P
Patricia GreenSep 6, 2025
What does 'Clr Active Memory?' mean on Siemens Measuring Instruments?
R
richard28Sep 6, 2025
If Siemens Measuring Instruments displays 'Clr Active Memory?', it's a response to pressing and holding the F4 key during power-up. Use the F4 key function to restore operation if a severe event disrupts system operation.
M
martin13Sep 8, 2025
What happens when Siemens Measuring Instruments asks 'Clr Saved Data?'?
J
joanrobinsonSep 8, 2025
If Siemens Measuring Instruments displays 'Clr Saved Data?', it only appears after pressing the in response to [Clr Active Memory?]. Answering Yes to [Clr Saved Data?] will erase ALL saved data. To invoke in RS-232 serial mode, type @@@ and then press key.