Connecting
5.5 Sensor Installation
FUS1010 IP65 NEMA 4X & IP66 NEMA 7
Operating Instructions, 01/2013, A5E02951520-AC
57
3. After receiving the spacing dimensions from the Installation Menu, prepare the pipe
surface. De-grease the surface, if necessary, and remove any grit, corrosion, rust, loose
paint, etc. Use abrasive material provided to provide a clean contact surface for the
sensors.
Note
Please note that the instructions show vertical mounting for clarity purposes only. Do not
install sensors on the top of a pipe.
Pipe
Cleaned Areas
Figure 5-9 Pipe Surface Preparation
4. Clean an area 13 mm (1/2-inch) on either side of the sensors.
5. Clean an additional 13 mm (1/2-inch) along the length of the sensors.
5.5.2 Sensor Identification and Selection
Sensor identification
The sensor part number located on the front face provides a detailed identification. For
example, the Part Number: 1011HNS-D1T1-S2 means:
 '+ 87+6 61)
Model Size
Hi Precision Temperature
NEMA w/F-Conn Agency Approved
Pipe Material