Functions
7.4 Span Data
FUS1010 IP65 NEMA 4X & IP66 NEMA 7
114 Operating Instructions, 01/2013, A5E02951520-AC
To change the default Span Data settings:
1. At [Meter Type], press <Right Arrow> to [Dual Path Flow] and press <ENTER>.
2. Scroll to [Span/Set/Cal] and press <Right Arrow>.
3. Highlight [Span Data] and press the <Right Arrow>.
6SDQ6HW&DO
6LHPHQV 'XDO3DWK$%&
6SDQDQG9DOXHVIRU$QDORJ'DWD
6SDQ'DWD
6HW$ODUP/HYHOV
&DOLEUDWH)ORZUDWH
,QWULQVLF
4. Highlight [Max Flow] and press <Right Arrow> to. Input 100% flow rate numeric data for
20mA. Press <ENTER> to store data.
5. Scroll down to [Min Flow].Press <Right Arrow> to input 0% flow rate numeric data for
4mA. Press <ENTER> to store data.
6SDQ'DWD
6LHPHQV 'XDO3DWK$%&
6HWP$!IORZUDWH
0D[)ORZ*$/0,1
0LQ)ORZ*$/0,1
0D[9V06
0LQ9V06
① Input numeric flow data here