22. As demonstrated in Figure 5-5, attach the connector of the A70-2 PPCL
PCB to the A70 DVSC PCB located in the host module enclosure.
23. Install the four M3 X 6 (CK3 X 6) screw, supplied in the kit, through the A70-
2 PPCL PCB, into the A70 DVSC PCB as indicated in Figure 5-5.
24. Place the PPCL shielding plate on the host module, in its original position
(Figure 5-2).
25. As depicted in Figure 5-6, install the four screws that were removed from
the PPCL shielding plate shown in Figure 5-2.
26. Install connector J2 in the PPCL shielding plate and secure it in place with
the two M2.5 X 6 screws and external lock washers that are supplied in the
kit.
Figure 5-5 Installing the A70-2 PPCL PCB
!
Caution: Use care to avoid dropping loose hardware into the system during the
disassembly and assembly of the G50. Loose hardware can short electronic
components. Safeguard all hardware for reinstallation.