Siemens 7482255 Rev 01 Page 25 of 52
October 2002
Chapter 4 - Removing and Replacing Spare Parts
6. Release the cables from the cable clamp located at the upper left of the
DIMAQ-IP module.
7. Remove the cable clamp located at the right side of the DIMAQ-IP; this
cable clamp is shown on the right side in Figure 4-20 and requires the
removal of the screw.
8. Disconnect the three cables from DIMAQ-IP module: When disconnecting
the DIMAQ-IP power cable, unscrew its cable clamp as one piece
(Figure 4-20).
9. Disconnect the remaining cables from the DIMAQ-IP module.
10. Loosen the three thumbscrews of the DIMAQ-IP module shown in
Figure 4-21, using the short number 2 Phillips screwdriver.
11. Create sufficient clearance for dismounting the DIMAQ-IP module safely by
routing the cables in front of it, and to the sides of the DIMAQ-IP module.
12. Move the DIMAQ-IP module to the left and dismount it from the system.
Figure 4-20 DIMAQ-IP Module Installation
Tie-wrap these cables to the adjacent panel.
DIMAQ-IP
!
Caution: The close clearances around the DIMAQ-IP create the potential for
cable damage. Verify that no cable is pinched or abraded when the DIMAQ-IP is
removed.