Chapter 4 - Removing and Replacing Spare Parts
2.0 Removing and Installing the Monitor
Use the procedures provided in this section to remove and to install the color
monitor on the G50 and the G60 S. Replacement of the monitor requires
approximately 15 minutes.
2.1 Tools Required
The following tools are required to perform this procedure:
One number 2 Phillips screwdriver
One 0.25 inch blade screwdriver
One Allen wrench (supplied with the system)
2.2 Removing the Monitor
1. Power down the system and unplug the power cable from the outlet.
2. As shown in Figure 4-3, remove the two cap screws from the top of the rear
OEM bay.
3. Remove the two screws that secure the rear upper panel and remove the
panel.
4. Remove two screws to remove the monitor post cover shown in Figure 4-4.
5. Remove two screws to remove the monitor cover shown in Figure 4-4.
!
Warning: Lifting weights of 15.9 kg (35 lb) or greater, or lifting in an awkward position can
cause serious injury. The monitor of the weighs 18 kg (40 lb); do not attempt to lift it without
help; this monitor is a two person lift. Use two people and proper lifting methods to avoid seri-
ous muscle and back injuries.
!
Warning: The G50 and the G60 S contains hazardous voltages that can range from 90 Vac to
264 Vac. These voltages can cause serious injury or death. Only authorized service personnel
are permitted to remove protective covers and access the energized components of this sys-
tem.