Page 4 of 16 7482255 Rev 01 Siemens
October 2002
Installing the G50 DIMAQ-IP
7. Loosen the two screws located at the bottom of the parallel PC link (PPCL)
shielding plate and remove the plate (Figure 5-2).
8. Use a pair of side cutters to cut the fastener attached to the USB cable,
above the RFC-5 noise filter (Figure 5-2).
9. Use the straight-blade screwdriver to unlock the two latches that hold the
RFC-5 noise filter to the USB cable (Figure 5-2) and remove the noise filter
from the cable.
10. Remove the cable clamp from the USB cable and disconnect the USB
cable from the PPCL shielding plate (Figure 5-2); the shielding plate should
now be completely separated from the system.
11. Place the PPCL shielding plate on a work surface as shown in Figure 5-3,
with the inside of the plate facing upward: The three wire ties supplied with
the kit will be installed on the plate to secure the USB cable.
12. Use the diagram shown in Figure 5-3 to locate the IN and OUT holes for A,
B, and C on the shielding plate: Count the holes in the perforated panel to
determine the exaction installation location and mark the IN and OUT
holes.
13. Insert, but do not fasten, one of the 3 wire ties supplied with the kit, at posi-
tion A-IN, B-IN, and C-IN of the PPCL shielding plate: The serrated surface
of the wire tie must be positioned on the inside of the tie when the tie is fas-
tened.
14. Install each wire tie through the IN hole and out of the OUT hole indicated,
at the exact location shown.
!
Caution: When cutting wire ties and fasteners, use care to avoid damaging the
cables or the associated equipment.
Figure 5-2 Removing the PPCL Shielding Plate