Page 40 of 52 7482255 Rev 01 Siemens
October 2002
Removing and Installing the MO Drive
8.3 Installing the MO Drive
Before installing the MO drive, move the number 2 and the number 4 DIP
switches to the On position as shown in Figure 4-36.
To install the MO drive, perform the procedures, in reverse sequence, that are
described in "Removing the MO Drive" (page 4-36).
8.4 Confirming the Operation of the MO Drive
1. Connect the system power cable to a properly grounded outlet.
2. Connect a transducer to the G50 or the G60 S.
3. Turn the power switch On.
4. Turn the Standby switch On.
5. Confirm that a B-mode image is displayed on the monitor.
6. Insert an MO disk into the MO drive: Verify that the MO disk is not write-pro-
tected.
7. Press FREEZE.
8. Save the frozen image.
9. Load the saved image from the MO disk and verify that the image displays
properly
!
Warning: Use threaded fasteners of the length cited here to avoid damaging or shorting under-
lying equipment.
Figure 4-36 MO Drive Top View: DIP Switch Configuration