EasyManua.ls Logo

Siemens SONOLINE G50 - Installing the Power Module; Right Side (RH) and Top View of the Power Module

Siemens SONOLINE G50
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 48 of 52 7482255 Rev 01 Siemens
October 2002
Removing and Installing the Power Module
9. Disconnect the cable from the connector P1of A07 PC board.
10. Disconnect the cables from the connector J1 and J2 of A03 PC board
(Figure 4-43).
11. Push and lift the power module to the right; then lift and pull the power mod-
ule out of the system.
11.3 Installing the Power Module
Use the procedures, in reverse order, that were described in "Removing the
Power Module" (page 4-47) to install the power module.
Figure 4-43 Right Side (RH) and Top View of the Power Module
Note: If the power module cannot be easily removed, loosen the two screws
located on the side panels, and move the panels to obtain additional clearance.
!
Caution: Before installing the power module, move all cables aside to prevent
damaging them.
!
Caution: Use fasteners of the correct length to prevent damage to underlying
equipment. See "Fastener Reference" (page H-1).
!
Caution: Figure 4-43 shows only J1 and J2. There is a J3 connector not indi-
cated in the photo. The J2 cable could be inadvertently installed in the J3 con-
nector. Verify that the correct cable is installed at the J2 connector.

Table of Contents

Other manuals for Siemens SONOLINE G50

Related product manuals