October 2002
Installing the G50 DIMAQ-IP
42. Attach the DIMAQ-IP option label at the designated option-label location.
1.4 Testing the DIMAQ-IP
Use the following steps to test the DIMAQ-IP module.
1. Connect the G50 power cord to the outlet, and energize the G50 system.
2. Verify that the G50 initializes without displaying error messages.
3. Press the Review key.
4. Verify that the DIMAQ-IP Digital Lab screen is displayed.
5. Verify that the DICOM Screen menu item is listed on the Change Screen
menu located at the lower-left side of the screen: This indicates that the
DIMAQ-IP software is version 1.2 or greater.
6. Operate the trackball and verify that the pointer moves across the screen.
7. Use the Digital Lab features to process several studies.
1.5 Removing the DIMAQ-IP Module
See "Removing and Installing the DIMAQ-IP Module" (page 4-24) for remove
and replace procedures appropriate for both the G50 and the G60 S DIMAQ-IP
module.
Figure 5-10 Installing DIMAQ-IP and Host Module Cables
*1
*2
*3
*4
*5
*5: VGA cable
Host Module