Chapter 6 - Troubleshooting
Figure 6-7 Chart C: B-mode Imaging
C
B-mode image
Ok?
Perform loop test 4.
Service >
Diagnostics >
E Module Package >
Pattern Check.
Recycle power.
Replace host module
Replace G60 S E module.
Replace power module.
Replace E module.
Replace E module.
TX power
adjust Ok?
CINE mode
Ok?
All B-mode
problems
resolved?
Probe select
Ok?
No
No
Yes
Yes
Yes
Yes
Return
Yes
Failed loop test 4;
passed loop test 3.
Replace cable
Replace E module
No
No;
4 pass
3 fail
Yes;
4 fail
3 pass
Perform loop test 3.
Service >
Diagnostic >
Power Module.
Loop test 4 and
3 both Ok?
Yes
No
Test; Service >
Diagnostics >
Power Module.
After test pattern
check, recycle the
system power
Power module
diagnostic
passed.
No
No
Replace E module.Yes
G60 S SystemNo
Yes
Third Array
Port Option
Installed
No Replace G50 LC E moduleNo
Remove Third
Array Port and
Retest
Yes
Problem
Remains
Replace G50 LC E module
Replace G50 Third Array Port
Yes
No