Page 4 of 52 7482255 Rev 01 Siemens
October 2002
Removing and Installing the Monitor
6. Remove two screws to remove the bottom cover shown in Figure 4-4.
7. Disconnect the four cables shown in Figure 4-5: two control cables, 1
power cable, 1 safety ground cable.
Figure 4-3 Host Module Cover
Figure 4-4 Monitor and Post Covers
!
Warning: The improper connection of the safety ground can create a shock hazard. The G50
and the G60 S contain hazardous voltages that can range from 90 Vac to 264 Vac. These volt-
ages can cause serious injury or death. Verify that the safety ground is attached using the cor-
rect hardware (see Figure 4-5) and that the screw is tight.