Siemens 7482255 Rev 01 Page 3 of 22
October 2002
Chapter 6 - Troubleshooting
15. Select Color Print for Set Up for PRINT/STORE 2 key and press SET.
16. Select Save and press SET.
17. Press the Menu key on the CP900 remote control.
18. Select Signal Set and set the input to RGB.
19. Press PRINT 2.
20. Confirm that the B-mode image on the monitor is printed correctly on the
CP900 printer.
21. Turn on the power switch of SVO-9500 VCR and insert a S-VHS tape into
the VCR.
22. Press F4 to display the Menu screen on the monitor.
23. Move the trackball upward and select Peripheral from the submenu.
24. Press SET.
25. Select S-VHS for Video Input Source and press SET.
26. Select VCR for the External RS-232C Port 1 and press SET.
27. Select Save and press SET.
28. Press the D key twice to invoke 2D + pwD mode.
29. Press RECORD to record the image and sound.
30. After pressing VIDEO I/O, press REWIND on the LCD, then press PLAY to
review the image and the sound just recorded.
31. Confirm that no error message is displayed on the monitor.
1.3 Verifying the Operation of the DIMAQ-IP Module
1. Press REVIEW.
2. Confirm that the monitor shows DIMAQ-IP module (Digital Lab) screen.
3. Move the trackball: Confirm that the cursor follows the movement of track-
ball.
2.0 Diagnostic Messages
Table 6-1, “Diagnostic Messages,” on page 6-4 lists the diagnostic messages
and the error messages likely to be encountered when a failure occurs on a
specific printed circuit board (PCB).