Diagnostic Messages
Page 4 of 22 7482255 Rev 01 Siemens
October 2002
Table 6-1 Diagnostic Messages (Sheet 1 of 5)
No. Diagnostics Message Error Log Message Expected
Failure Location
Failed
PCB
Failed
PCB
Name
1 Fail : Operator: add=0xXXXX val(0x55)=0xXX Fail : Operator: add=0xXXXX val(0x55)=0xXX Control Panel A10 OPCT
2 Fail : Operator: add=0xXXXX val(0xaa)=0xXX Fail : Operator: add=0xXXXX val(0xaa)=0xXX Control Panel A10 OPCT
3 Fail : HostBox : HostBox Return Path(Card-Host Test Pat.) Fail : HostBox : HostBox Return Path(Card-Host Test Pat.) Host Module A70 DVSC
4 Fail : CardBox : CardBox Return Path(Card-Host Test Pat.) Fail : CardBox : CardBox Return Path(Card-Host Test Pat.) E Module - E Module
5 Fail : HostBox : HostBox Test Pat. Gen.(Card-Host Test Pat.) Fail : HostBox : HostBox Test Pat. Gen.(Card-Host Test Pat.) Host Module A70 DVSC
6 Fail : CardBox : CardBox(Card-Host Test Pat.) Fail : CardBox : CardBox(Card-Host Test Pat.) E Module A66 DTPS
7 Fail :Card/Host: CardBox and HostBox(Card-Host Test Pat.) Fail :Card/Host: CardBox and HostBox(Card-Host Test Pat.) E Module or Host
Module
A66/A70 DTPS/
DVSC
8 Fail : Cable : Cable between Card and Host(Card-Host Test
Pat.)
Fail : Cable : Cable between Card and Host(Card-Host Test
Pat.)
Cable between E
Module and Host
Module
-Cable
9 Fail : HostBox : HostBox(Card-Host Test Pat.) Fail : HostBox : HostBox(Card-Host Test Pat.) Host Module A70 DVSC
10 Fail : HostBox : HostBox Return Path(DIMAQ-IP-Host Test
Pat.)
Fail : HostBox : HostBox Return Path(DIMAQ-IP-Host Test
Pat.)
Host Module A70 DVSC
11 Fail : DIMAQ-IP : DIMAQ-IP Return Path(DIMAQ-IP-Host
Test Pat.)
Fail : DIMAQ-IP : DIMAQ-IP Return Path(DIMAQ-IP-Host
Test Pat.)
DIMAQ-IP A75 SPCL
12 Fail : HostBox : HostBox Test Pat. Gen.(DIMAQ-IP-Host Test
Pat.)
Fail : HostBox : HostBox Test Pat. Gen.(DIMAQ-IP-Host Test
Pat.)
Host Module A70 DVSC
13 Fail : DIMAQ-IP : DIMAQ-IP(DIMAQ-IP-Host Test Pat.) Fail : DIMAQ-IP : DIMAQ-IP(DIMAQ-IP-Host Test Pat.) DIMAQ-IP A75 SPCL
14 Fail :DIMAQ-IP/Host: DIMAQ-IP and HostBox(DIMAQ-IP-
Host Test Pat.)
Fail :DIMAQ-IP/Host: DIMAQ-IP and HostBox(DIMAQ-IP-Host
Test Pat.)
DIMAQ-IP or
Host Module
A70/A75 DVSC/
SPCL