Siemens 7482255 Rev 01 Page 3 of 8
October 2002
Chapter 7 - Installing the Version 1.2 Software Update on the G60 S
1.1.2 Backing Up Presets
Use the following procedure to back up all presets, QuickSets, OB tables, and
formulas
1. Press the Review button to view the Study Screen.
2. Select the MO Disk radio button.
3. Select the Preset tab.
4. Select the Store Quicksets button.
5. Check the All Presets checkbox and select the Ok button.
6. In the File Type menu, select General System Presets.
7. Select the Store button.
8. Enter a file name using the keyboard: Include the date and data description
in the file name (MMDDYYGenSysPre).
9. Select the Ok button; select the Ok button to close the message box and to
continue.
10. In the File Type menu, select OB Tables and Formulas.
11. Select the Store button.
12. Enter a file name using the keyboard: Include the date and data description
in the file name (MMDDYYOBTabFor).
13. Select the Ok button; select the Ok button to close the message box and to
continue.
14. In the File Type menu, select Default Settings.
15. Select the Store button.
16. Enter a file name using the keyboard: Include the date and data description
in the file name (MMDDYYDefSet).
17. Select the Ok button; select the Ok button to close the message box and to
continue.
18. In the File Type menu, select Total System Configuration.
19. Select the Store button.
20. Enter a file name using the keyboard: Include the date and data description
in the file name (MMDDYYTotSysCon).
21. Select the Ok button; select the Ok button to close the message box and to
continue.
22. Select the Utility tab and double-click MO and SONOLINE to view the con-
tents of the MO diskette.
Note: Refer to the “Documentation and Storage” chapter of the System Refer-
ence manual for the preset backup procedure.