8. Use the Allen wrench to remove the two socket head cap screws and
remove the single bolt shown in the upper right corner of Figure 4-6: Hold
the monitor securely while removing these screws.
9. Remove the monitor from the post.
10. Remove the mounting hardware from the original monitor; this hardware
will be installed on the replacement monitor.
Figure 4-5 Attaching the Monitor Cables
Safety Ground
Screws identified as CK4 X 12 in Figure 4-5 are M4 X 12 binding head with external
lock washer.
The safety ground cable is attached with an M4 X 12 screw through the lug, with an
external toothed M4 lock washer located between the lug and the chassis.