Siemens 7482255 Rev 01 Page 15 of 16
October 2002
Chapter 5 - Installing Equipment Options
10. Verify that an assembly number is displayed in the AS No. field for the A36-
2 PCB (Figure 5-15).
11. Select the Home button to return to the Main Menu screen.
12. Select the Exit button exit the Service software; the ultrasound imaging
screen will be displayed.
13. Acquire an ultrasound image, using the probe attached to the new port.
14. Verify that no vertical line or image noise is present on the acquired image.
15. Verify that the option label for the new port is attached at the rear of the sys-
tem.
16. The new port is now properly installed and operational.
Note: If the A36 A3RD-2 PCB is not installed in the G50, the screen shown in
Figure 5-16 will display the message “Not Installed” in the AS No. column for the
A36-2 PCB on the System Configuration – Hardware screen.
Figure 5-15 System Configuration – Hardware Screen (A36-2 Installed)
Home
Hardware
Software
Peripheral
System Configuration - Hardware
Servi ce
Site
System
HW/SW
Assy No. Assy Name AS No.
A10 OPCT 0
A30 ECGX 101
A36 PSEL Not Install ed
A36- 2 A3RD 2
A40 PAMP Not Install ed
A40- 1 CWRX Not Install ed
A40- 3 RX32 Not Install ed
If A36 A3RD-2 PCB has been
successfully installed, a numeric
value is shown here.
XXXX