Page 2 of 22 7482255 Rev 01 Siemens
October 2002
Verifying the System Operation
1.0 Verifying the System Operation
Use the following methods to verify that the G50 or the G60 S is operating
properly.
See Instructions for Use for detailed operating instructions.
1.1 Verifying the Operation of the System
1. Connect the power cable to the wall outlet.
2. Connect a transducer to the ultrasound system.
3. Turn the power switch On.
4. Turn the Standby switch On.
5. Confirm that a B-mode image is displayed on the monitor.
6. Verify that a Color flow image is displayed on the monitor.
7. Verify that the Doppler spectrum is displayed on the monitor.
8. Verify that the ECG trace is displayed on the monitor.
1.2 Verifying the Operation of the Peripherals
Use the following instructions to verify the operation of peripherals attached to
the G50 or the G60 S.
1. Turn on the power switch of the P91W printer.
2. Press F4 to display the Menu screen on the monitor.
3. Move the trackball upward and select Customize Keys from the left frame
displayed on the screen.
4. Press SET.
5. In the right frame of the displayed screen, select B/W Print for
PRINT/STORE 1 key.
6. Move the cursor to the radio button labeled B/W Print, and press SET.
7. Select B/W Printer for Set Up for PRINT/STORE 1 key and press SET.
8. Select Save and press SET.
9. Press PRINT/STORE 1.
10. Confirm that the B-mode image on the monitor is printed correctly on the
P91W printer.
11. Turn on the power switch of the CP900 color printer.
12. Press F4 to display the Menu screen on the monitor.
13. Move the trackball upward and select Customize Keys from the submenu.
14. Press SET.