Page 6 of 8 7482255 Rev 01 Siemens
October 2002
Installing the Version 1.2 Software Update
38. Reload the image file from the CD to the system.
1.3 Installing the Version 1.2 System Software Update
Use the following steps to install the version 1.2 system software update.
1. Insert the MO diskette containing the version 1.2 software in the MO drive.
2. Press the F4 key to display the Preset screen.
3. Use the trackball to scroll down the menu to the System/Language menu
item and select that menu item.
4. Press SET.
5. Select the Retain Current Preset radio button: the system automatically
stores the settings for all existing system presets, QuickSets, OB tables
and formulas, and the current language selection.
6. Select the Upgrade System button.
7. The system instructs the user to restart the system: Restart the system by
pressing the green On/Off switch to de-energize the system.
8. Wait 15 seconds and then press the green On/Off switch to energize the
system.
9. The system will display a progress bar at the bottom of the screen, as the
system software installation proceeds; the loading of the software requires
approximately 10 minutes.
10. When the software installation process is completed, the system displays
the active image screen.
11. Press the MO eject button to remove the diskette.
1.4 Restoring the Service Data
Use the following instructions to restore the service data.
1. Press the F4 key and select the Service menu item.
2. Enter the service password and select the Ok button to display the Main
Menu screen.
3. Select the Backup/Restore button to display the Backup / Restore screen.
4. Select the Restore radio button located at the top of the screen.
5. Check the All Packages checkbox to restore all service data.
!
Caution: Do not attempt to reload the DIMAQ-IP version 1.023 software to a
system that is operating with version 1.2 system software installed. This opera-
tion will fail and display the following message: “Please check the DIMAQ-IP
software version; DIMAQ- IP will shut down soon.”
Note: If additional disks are required to complete the installation, follow the
instructions displayed on the screen.