Page 42 of 52 7482255 Rev 01 Siemens
October 2002
Removing and Installing the Host Module
7. Disconnect the cable from the connector J36.
8. Remove four screws from the bottom of the host module.
9. Disconnect all of the cables from the upper side of the host module.
10. Disconnect all of the cables from the OEM subpanel located on the left side
of the host module.
11. Disconnect all the cables from the OEM interface panel shown in Figure 4-
38.
12. Loosen the four cap screws located at the rear of the host module.
13. Remove the host module from the G50 or the G60 S.
9.3 Installing the Host Module
1. Verify that the cables are securely positioned in both slits located at the
sides of the opening shown in Figure 4-39.
Figure 4-38 OEM Interface Panel
Figure 4-39 Routing the Host Module Cable