Page 34 of 52 7482255 Rev 01 Siemens
October 2002
Removing and Installing the Keyboard
5. Disconnect the cable from connector J3 of the A10 PC board shown in
Figure 4-28.
6. Unlatch the cable clamp attached to the base of the upper frame shown in
Figure 4-29.
7. Holding the keyboard, remove the two screws from each of the two metal
slides: four screws total.
8. Pull the keyboard slowly toward you to remove it from the upper G50 or the
G60 S.
7.3 Installing the Keyboard
Install the keyboard by following the instructions, in reverse sequence,
provided in "Removing the Keyboard" (page 4-33).
Figure 4-29 Cables and Clamp Under the Control Panel
Safety Ground
Safety Ground
The safety ground cable is attached with an M4 X 12 binding head screw through
the lug, with an external toothed M4 lock washer located between the lug and the
chassis.
!
Caution: Route the cable so that it does not obstruct the latching of the key-
board.
!
Caution: Use fasteners of the correct length to prevent damage to underlying
equipment. See "Fastener Reference" (page H-1).
!
Warning: The improper connection of the safety ground can create a shock hazard. The G50
and the G60 S contain hazardous voltages that can range from 90 Vac to 264 Vac. These volt-
ages can cause serious injury or death. Verify that the safety ground is attached using the cor-
rect hardware and that the installation screw is tight.