Siemens 7482255 Rev 01 Page 5 of 16
October 2002
Chapter 5 - Installing Equipment Options
15. Measure 260mm from the termination of the USB cable as shown in
Figure 5-4, and mark the cable.
16. Place the PPCL shielding plate adjacent to the G50 and align the 260mm
mark on the USB cable, with the wire tie designated A.
17. Loop the cable as shown in Figure 5-4, through wire tie B.
18. Align the loop so that the USB connector is adjacent to wire tie C.
19. Fasten wire tie B, realign the cable at A and C, and fasten the wire ties A
and C.
20. Verify that no portion of the cable or connector extends more than 9mm
from the PPCL shielding plate; extending beyond 9mm, the cable or con-
nector will interfere with the adjacent equipment.
21. After fastening the wire ties, remove the excess portion of the wire ties with
the side cutters.
Figure 5-3 Installing the Wire Ties
Figure 5-4 Measuring the USB Cable