Appendix B - Installation Report
Appendix B: Installation Report
Consult the Siemens Speed Info 001/2001/US for information on completing
the Ultrasound System Installation Report (7295749) shown in Figure B-1.
Access Speed Infos on the intranet at: http://www-td.med.siemens.de:88/.
From this intranet page, select the following menu items to view SI 001/2001/
US:
Product Information → US Systems → Speed Infos US → 2001 → 001/2001/US
During the installation, perform the Field Electrical Safety Test Procedure (p/n
7482776) using the G50 and G60 S Electrical Safety Test Limits (p/n
7482784); record the results of the safety test in the Field Electrical Safety Test
Report (p/n 7482768).
After completion of the electrical safety test and the test report, provide the
customer and the regional office with a copy of the report; retain a copy of the
report for your personal records.
Note: The Ultrasound System Installation Report presented in Figure B-1
includes instructions for its completion and submission. Submit the report within
5 working days after completing the installation activity.