Siemens 7482255 Rev 01 Page 21 of 52
October 2002
Chapter 4 - Removing and Replacing Spare Parts
27. Using the diagrams provided in Figure 4-18 and Figure 4-19, connect the
cables between the rear side of the SVO-9500 and the interface panel and
between the rear side of the P91W and the interface panel.
28. Secure the OEM bay to the system, using the four M4 X 8 binding head
screws with external lock washers that were removed.
29. Re-install the top cover on the G50 or the G60 S.
30. Re-install the front cover on the G50 or the G60 S.
Figure 4-18 SVO-9500 VCR Connections
Figure 4-19 SVO-9500 and Interface Panel