Siemens 7482255 Rev 01 Page 5 of 22
October 2002
Chapter 6 - Troubleshooting
15 Fail : Cable : Cable between DIMAQ-IP and Host(DIMAQ-IP-
Host Test Pat.)
Fail : Cable : Cable between DIMAQ-IP and Host(DIMAQ-IP-
Host Test Pat.)
Cable between E
Module and
DIMAQ-IP
-Cable
16 Fail : HostBox : HostBox(DIMAQ-IP-Host Test Pat.) Fail : HostBox : HostBox(DIMAQ-IP-Host Test Pat.) Host Module A70 DVSC
17 Fail : HostBox : DIMAQ-IP-HostBox Test Fail : HostBox : DIMAQ-IP-HostBox Test Host Module A70 DVSC
18 Fail : HostBox : CardBox-HostBox Test Fail : HostBox : CardBox-HostBox Test Host Module A70 DVSC
19 Work Memory Error! - Recycle power.
If the error mes-
sage is displayed
again, exchange
Host Module.
-/A70 -/DVSC
20 Fail : D+3V A+3V=Upper Voltage Error! : V=XXXX etc Power Module A03/A07 SWRG/
PDST
21 Fail : A-10V A-10V=Lower Voltage Error! : V=XXXX etc Power Module
22 Fail : A+5V A+5V=Lower Voltage Error! : V=XXXX etc Power Module A03/A07 SWRG/
PDST
23 Fail : D+5V D+5V=Lower Voltage Error! : V=XXXX etc Power Module A03/A07 SWRG/
PDST
24 Fail : A-12V A-12V=Lower Voltage Error! : V=XXXX etc Power Module A03/A07 SWRG/
PDST
25 Fail : A+12V A+12V=Lower Voltage Error! : V=XXXX etc Power Module A03/A07 SWRG/
PDST
26 Fail : A+90V A+90V=Lower Voltage Error! : V=XXXX etc Power Module A66 DTPS
Table 6-1 Diagnostic Messages (Sheet 2 of 5)
No. Diagnostics Message Error Log Message Expected
Failure Location