EasyManua.ls Logo

Siemens SONOLINE G50 - Page 130

Siemens SONOLINE G50
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 20 of 22 7482255 Rev 01 Siemens
October 2002
Diagnostic Flow Charts
Figure 6-10 Chart F: ECG
F
ECG trace
display Ok?
ECG gain
adjustment
Ok?
ECG display
offset Ok?
ECG trigger
mode Ok?
Doppler trace
display Ok?
ECG trace has
red marker?
Return
Replace E module.
Replace ECG module.
Replace E module.
Replace ECG module.
Replace ECG module.
Replace E module.
Yes
Yes
Yes
Yes
No
No
No
Yes
No
No No
Yes

Table of Contents

Other manuals for Siemens SONOLINE G50