Ionics Instruments 2004 Page 11-7 DLM 30007-08 Rev. A
GD950302 08:00 950808 13:30
indicates that the first data on the card is from March 2, 1995 at 8:00 AM and
the last data is from August 8, 1995 at 1:30pm.
To download all or part of the data using the GD command using the same
format. For example:
GD950701 00:00 950702 00:00
will retrieve all of the TOC data from midnight July 1 to midnight July 2, 1995.
The following format must be used:
GDYYMMDD (space) HH:MM (space) YYMMDD (space) HH:MM (enter)
The backspace or delete key cannot be used to correct mistakes. If a mistake
is made, simply hit the enter key and retype the command. If the format is not
correct, the analyzer will respond with IV (invalid command) or an ER (error).
For a valid command such as gd950808 13:37 950808 14:01, the analyzer
will respond with:
CHANNEL NAMES:STRM NAME 1,STRM NAME 2,STRM NAME 3, STRM
NAME 4, STRM NAME 5,STRM NAME 6,STRM NAME 7,STRM NAME 8
The first two line are a header that contain the stream names used in the
multistream sampler. If not using the multistream samples, the header is not
used.