Ionics Instruments 2004 Page 11-8 DLM 30007-08 Rev. A
The next line contain the first sample data. The format is:
YYMMDDHH:MM TOC(units) IC(units) (sample mode) (sample number)
The units are either B for ppb or M for ppm and the sample mode is blank for
on-line, G for grab samples, T for tubes (autosampler), S for stream
(multistream sampler) and F for Turbo mode.
95080813:27 22.8B 55.2B - On-Line
95080813:43 16.1B 37.1BG1 - Grab
95080813:49 19.3B 45.1BG2 - Grab
95080813:55 14.1B 30.9BT1 - Tubes (autosampler)
95080814:01 12.1B 26.1BS1 - Multistream (number one, in this case)
95080814:09 12.9M 29.3M F - Turbo
For this example, the first sample data is from August 8, 1995 at 1:37pm the
TOC was 22.8 ppb and the IC was 55.2 ppb. The sample mode is blank,
designating on-line measurements. Successive measurements are list on new
lines.
An example of the reply for the grab sample mode is 95080813:43 19.3B
45.1BG1, where the G designates the grab sample mode and the 1 indicates
that this was the second sample run this day.
Depending on the communication program, save the reply as a text file or
copy the data to a text editor or spreadsheet program.