11.12 WATER CONDUCTIVITY – ID.1466
Conductivity probe management is performed only if the “IC6010” optional is correctly installed.
Required CPU Master firmare version 7.16.0 (july 2017) or subsequents.
The probe present flag is factory set; the operator – if he has a superuser password – can:
1. enable/disable the probe;
2. set the check threshold value.
A single threshold is used for all programs which include a conductivity check.
The maximum possible value of the threshold is 50µS/cm. This is the full range value of the system installed on the
machine. If the conductivity reading is higher than 50 µS, the actual value is not displayed, but the generic
value “K > 50” displays.
Measure the conductivity of the demineralised water in your system before using it to ensure
you have set an appropriate threshold value.
The machine's threshold value may not be lower than the conductivity of the demineralised
water in your system.
Check the water supply from time to time to check that it maintains its original
characteristics.