6.4 ACCESSING AND REUSING THE DEVICE AFTER AN INCOMPLETE CYCLE
Instructions on the safety of the device in case of incomplete operation are given in compliance with the
par. 5.4.4.g and 13.1.102 IEC61010-2-040:2015.
In particular, these precautions are applied in the event the device is used to treat biologically
contaminated material.
WARNING
If a disinfection cycle is interrupted (by the operator or if an alarm arises): pay attention in
handling all instruments of the load inside the washing chamber.
The loaded items and internal parts of the machine might be infected/ biologically
contaminated.
Before handling the instruments or before any maintenance operation: start a complete cycle
of thermal disinfection or, if it is not possible to start a complete cycle, handle the instruments
cautiously (using the protection devices suitable for handling infected instruments, e.g. gloves,
smock).
WARM SURFACE DANGER
The machine operates a cycle of thermal disinfection using high temperature water, up to 93°C,
and detergents. If, in case of malfunction, the door is open and there is water inside the tank:
avoid skin contact, for burning and irritation danger due to the toxicity of chemical products.
Never touch the heating elements in the tank.
Contact authorized technicians in case of malfunction.
6.5 USE OF THE RELEASE TROLLEY
Sliding door devices: The use of the original release trolley is necessary for the safe transport the
washing cart inside and outside the machine.
Contact the authorized dealer for information about the most suitable multi-purpose trolley.
Sample image of the release trolley for washing cart transport and device tending.