Water conductivity - Scroll through the list of machine
parameters to select “Water conductivity”.
This displays screen 1466.
Enable - Touch the Enable box: the check mark indicates
whether the probe is enabled or not.
If the probe is disabled, the cycles run as though the probe is
not installed.
Threshold: Touch the “±” button to enable the +/- buttons for
adjusting the threshold value.
Confirm the setting with the button at the bottom right.
11.12.1 Purpose of the accessory
The conductivity probe, part code IC6010, is an optional kit for determining the conductivity of the rinse
water.
The conductivity value is used to determine whether further rinses are necessary, depending on whether
a given threshold has been satisfied or not.
The accessory must be ordered together with the instrument or glassware washer, and can only be
installed at the factory.
11.12.2 Operation of the machine with the probe active: conductivity check function
To use the conductivity check function, the following must obtain:
1. The probe must be installed;
2. The conductivity threshold must be set;
3. The program must include a conductivity check, identified on the display as “Cond. Check”.
The standard factory programs are configured to include a conductivity check in the last rinse phase.
Custom programs may be edited to include the conductivity check as necessary. In general, we recommend including
the check in the last phase of the program, using demineralised water.
For details on how to edit custom programs, refer to the device's user manual.
Operating mode during the cycle: the machine runs the conductivity check, in the appropriate phase, by comparing
the read value of K with the threshold value K_t (t=threshold).
- If K < K_t - the result is positive and the phase continues normally.
- if K ≥ K_t the chamber is emptied out and supplementary fill and rinse cycles are run (up to 3) until the read value falls
below the threshold.