The Superuser can modify the parameters available:
- Code validity timeout: the time in minutes
available from acquisition of the codes to
closing of the door. If the door is not closed
by the end of the timeout, the codes are
automatically reset and must be reacquired
(guideline time max 15 min.).
- Rack:
o The Flag enables/disables the need
for a rack code before the cycle can
be started.
o n_char: number of alphanumeric
characters in the rack code (max 14)
- User:
o The Flag enables/disables the need
for a user code before the cycle can
be started.
o n_char: number of alphanumeric
characters in the code (max 14)
- Automatic timeout: time after which the
screen automatically switches, once the
codes have been acquired.
o Flag: enables automatic switching to
the next screen.
o Sec: the seconds for which the
acquisition screen is retained before
switching to the next screen.
Default values for number of characters: 13 for both
user and rack codes.