2017/05/29 12:48:41 Cycle started
ID. 12
Operator 255
Cycle N. 59
. . . . . . . . . . . . . . . . . . . .
12:48:41 Start phase N. 1
WASH
12:48:41 Load CW / 23L
12:52:29 P1 = 117 ml / 27°C
12:52:29 TL 27.90°C Required: N.D.
12:52:29 Holding time: 2:01
12:52:29 Heating ON
12:54:31 Min-Max 27.90°C - 33.70°C
. . . . . . . . . . . . . . . . . . . .
12:55:25 Start phase N. 2
THERMAL DISINFECTION
12:55:25 Load CW / 23L / 34°C
12:58:38 P1 = 117 ml / 23°C
12:58:38 Heating ON
13:13:00 TL 93.10°C Required: 93 °C
13:13:00 Holding time: 5:01
13:14:20 Heating ON
13:18:02 Min-Max 93.10°C - 94.20°C
. . . . . . . . . . . . . . . . . . . .
13:18:59 Start phase N. 3
RINSE
13:18:59 Load HW / 23L / 91°C
13:21:52 P2 = 72 ml / 41°C
13:21:52 TL 41.60°C Required: N.D.
13:21:52 Holding time: 2:01
13:21:52 Heating ON
13:23:54 Min-Max 41.60°C - 46.60°C
. . . . . . . . . . . . . . . . . . . .
[…]
. . . . . . . . . . . . . . . . . . . .
13:35:19 Start phase N. 6
RINSE
13:35:19 Load DW / 23L / 48°C
13:38:24 Heating ON
K = 20 us/cm
K_t = 20 us/cm
Check KO
13:43:35 Heating ON
K = 18 us/cm
K_t = 20 us/cm
Check OK !
13:45:33 Heating ON
13:52:40 TL 75.10°C Required: 75 °C
13:52:40 Holding time: 1:01
13:53:42 Min-Max 75.10°C - 76.60°C
13:54:38 DRYING
TL1 = 75.80 TA1 = 66.80
14:24:42 Drying complete
. . . . . . . . . . . . . . . . . . . .
Successfully completed cycle
Ao 7579.70
Cycle started: date and time of cycle start.
ID.12 - the ID number of the program (“ID_prog” in the
program table).
Cycle N. 59 – 59th program run by the machine.
Each phase has a header line: Start phase N.
This is followed by a description of the phase, such as:
WASH.
Load CW / 23L Load cold water, 23 litres.
P1 = 117 ml / 27°C amount of detergent batched by
pump P1
TL 27.90°C Required: N.D. – temperature of probe TL
and phase temperature setpoint. “N.D.” stands for Not
Defined: Heating is timed if no temperature setpoint is
specified; the duration is equal to the indicated holding
time.
Holding time: 2:01 – duration of phase, 2 minutes.
Min-Max 27.90°C - 33.70°C the minimum and
maximum temperatures TL for the most recent holding
time.
The record TL 93.10°C Required: 93°C identifies the
start of the holding time.
The record Min-Max 93.10°C - 94.20°C identifies the
end of the holding and the minimum and maximum
temperatures reached by the system during that time.
Conductivity check reading
K = 20 us/cm
K_t = 20 us/cm
Check KO
K is the value read by the machine's conductivity
probe, in µS/cm.
K_t is the threshold setting (t=threshold)
If K<K_t, the result is positive (“Check OK !”),
if not, it is negative (“Check KO”).
If the result is negative, the water is drained out of the
chamber, reloaded and a new check is run:
K = 18 us/cm
K_t = 20 us/cm
Ao 7579 is the value of the parameter Ao, calculated
during the thermal disinfection phase.