EasyManua.ls Logo

Squid Ink CoPilot 500 - Touchscreen Calibration

Squid Ink CoPilot 500
152 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
94 Software Interface
2005995 Rev O – 210712
Calibrating the Touchscreen
If the buttons on the Touchscreen no longer activate the corresponding functions or are
unresponsive, recalibration may be necessary. Select this button to calibrate the
Touchscreen.
To Recalibrate the Touchscreen:
1. On the Settings Screen, select the Calibrate Screen button.
2. In the Calibrate Screen, select the Calibrate button.
3. In the Calibrate utility, touch the center of each small box in turn as they appear.
4. After the center box has been touched, a “Calibration Completedmessage is
displayed. The Controller returns to the Settings Screen.
5. Restart the Controller for calibration results to take effect.
NOTE: If the screen is out of calibration and cannot be recalibrated using the
Controller, select Reset Calibration on the Printer Configuration Tools tab in
Orion. Refer to the Orion Software User Guide for more information.

Table of Contents

Related product manuals