Objet1000 Plus User Guide 2 Safety
DOC-06810 Rev.B 2-7
WhenhandlingUV-curedmodelsthatmaynotbecompletelycuredonthe
surface,commonlatexglovesareadequate.
Topreventrespiratoryirritation,ventilateareaswhereModelandSupport
materialsareused.
CleanupModel-materialandSupport-materialspillswithdisposable
towelsorotherabsorbent,non-reusablematerial,suchassawdustor
activatedcharcoal.Rinsethespillareawithdenaturedorisopropylalcohol
(IPA),followedbysoapandwater.Disposeoftheabsorbentmaterialin
accordancewithlocalregulations.
Donotwashcontaminatedclothingathome;clothingshouldbe
professionallylaundered.
Disposeofcontaminatedshoes,beltsandotherleatheritemsin
accordancewithanyapplicableregulations.Absorbedprintingmaterial
mayre-exposetheuserwhentheseitemsareworn.
UV Lamps
UVlampsusedbytheprintertocureprintingmaterialscontainasmall
amountofmercury.Intheunlikelyeventoflampbreakage,avoidinhaling
mercuryvapor,andventilatetheroom.Ifthelampruptures(breaks)during
operation,leavetheroomandventilateitthoroughlyforabout30minutes.
Useprotectiveglovestopreventcontactwithmercuryandotherlamp
components.Carefullyremovespilledmercurywithamethodthatprevents
thegenerationofmercuryvapor,suchasasyringe,packingtapeorpaper.
Placethebrokenlamp,mercuryandcontaminatedmaterialsinanair-tight,
non-metalliccontainer.Disposeofthecontainerinaccordancewith
applicableregulations.