CHAPTER 10 Troubleshooting
10-32 Sysmex XE-5000 Instructions for Use
Revised June 2008
The following operation cycle is displayed on the Counter
screen.
TOTAL: Indicates the operation cycle of the equipment.
CBC: Indicates the operation cycles in the CBC
analysis mode.
DIFF: Indicates the operation cycles in the DIFF
analysis mode.
NRBC: Indicates the operation cycles in the NRBC
analysis mode.
RET: Indicates the operation cycles in the RET
analysis mode.
HPC: Indicates the operation cycles in the HPC
analysis mode.
FFS: Indicates the operation cycle count of the FFS
reagent pump after replacement of the FFS.
SHUT: Indicates the analysis cycle after shutdown
execution.
PIAS: Indicates the piercer operation cycle after
replacement of the piercer.
SRV: Indicates the analysis cycle after the SRV
cleaning.
FCM-MT: Indicates the operation cycles of the FCM
sheath motor.
RBC-MT: Indicates the operation cycles of the RBC
sheath motor.
WB-MT: Indicates the operation cycle of the whole
blood aspiration motor.
LASER: Indicates the laser oscillation cycle.
BF-MEAS: Indicates the operation cycles in the body fluid
analysis mode.(except background check)
BF-BCHK: Indicates the operation cycles in the body fluid
analysis mode.(background check only)
4. Select PIAS or SRV by moving the cursor with or
key on the Counter screen.
5. When you select “OK” on the function menu of the
Counter screen, then select “OK” again, the highlighted
parameter counter will be reset.
When you wish to cancel the reset operation, select
“Cancel” on the function menu and return to the Status
Menu screen.
Information
Only PIAS and SRV are selectable by the cursor, and
operation cycle can be reset.
The other parameters are for the reference.