CHAPTER 10 Troubleshooting
10-34 Sysmex XE-5000 Instructions for Use
Revised April 2007
3. Select Sampler on the function menu of the Test Menu
screen.
The Sampler Test screen will appear.
4. Select “1 Rack feed in” by moving the cursor with or
key on the Sampler Test screen.
5. Select “Execute” on the function menu of the Sampler
Test screen. Rack feed-in operation will start. Check the
sampler operation.
6. To cancel the sampler test, select “Cancel” on the
function menu, or press the RETURN key.
2. Rack Shift
Select this program to execute the Rack Shift operation test.
Set the rack in place before executing the test.
1. Set the rack in place in the sampler analysis line.
2. From the Main Menu screen on the Main Unit, select
“TEST on the function menu.
The Test Menu screen will appear.
3. Select Sampler on the function menu of the Test Menu
screen.
The Sampler Test screen will appear.
4. Select “2 Rack shift” by moving the cursor with or
key on the Sampler Test screen.
5. Select “Execute” on the function menu of the Sampler
Test screen. Rack shift operation will start. Check the
sampler operation.
6. To cancel the sampler test, select “Cancel” on the
function menu, or press the RETURN key.
3. Rack Feed Out
Select this program to execute the Rack Feed Out operation
test. Set the rack in place before executing the test.
1. Set the rack in place on the left rack pool.
2. From the Main Menu screen on the Main Unit, select
“TEST on the function menu.
The Test Menu screen will appear.
3. Select Sampler on the function menu of the Test Menu
screen.
The Sampler Test screen will appear.
4. Select “3. Rack feed out” by moving the cursor with or
key on the Sampler Test screen.