CHAPTER 11 Technical Information
11-22 Sysmex XE-5000 Instructions for Use
Revised April 2007 EU
IMI scattergrams
Two-dimensional distributions of the cell size and density of
the cell interior (such as nucleus size) are drawn via the RF/
DC detection method.
The IMI analysis channel is a channel used to discriminate
between immature and mature white blood cells.
6. NRBC analysis
NRBC analysis procedure
The NRBC analysis is performed to classify and count the
nucleated red blood cell population in the blood. The NRBC
analysis procedure is explained as follows:
Sample rotor valve
CELLPACK
Sample tube
(blood: 18 μL)
Reaction chamber
Optical detector
block
STROMATOLYSER-NR(L)
lyse reagent: 0.882 mL
STROMATOLYSER-NR(S)
dye solution: 18 μL